Model DI-730-P - OBSOLETE (Replaced by DI-4730)
8-channel Isolated High Voltage Printer Port Data Acquisition System
Replaced by DI-4730
- Includes a DI-730-P instrument, a power supply and cable, an NIST-traceable calibration certificate, a 25-pin parallel port cable, and software on CD
- Bundle and Save
- Measurement range of ±10mV to ±1000VDC (or peak AC) over six ranges
- 1000V input-to-output and channel-to-channel isolation
- Perfect for troubleshooting AC or DC motor drive systems. Measure field voltages and currents, control signals, RPM, and more.
- Also for electric motor/generator maintenance and troubleshooting in mills and locomotion
- DI-730s easily replace bulky chart recorders
- Replace entire racks of isolation amps, high & low voltage amps, and data acquisition subsystems with one, portable, lightweight, 9 × 7 × 1½ inch instrument
- Up to 150 kHz Sample Throughput Rate
- Unlock code required for sample rates over 240 Hz only when using WinDaq software
- 14-bit resolution
Description
DI-730 High Voltage Printer Port Data Acquisition System with Windaq software. Features 8 wide range differential analog input channels, 150,000 Samples per second throughput, 14-bit measurement resolution, and full scale ranges of ±10, ±100 mV, ±1, ±10, ±100, ±1000 V full scale.
The DI-730-P offers 14-bit resolution and 150,000 Samples per second waveform recording capability. Use the DI-730 for wide dynamic range measurements, especially those involving AC or DC electric motors. Examples include paper, aluminum, and steel mills; high speed trains; and rail and seagoing locomotion. The high sample rate, exceptional isolation, and CMR characteristics make it well-suited for maintenance and troubleshooting of DC drive systems. Other applications include RPM measurements from motor/generators, supply voltage and current measurements, and field current measurements from drive-roll, braking, and take-up motors.
Accessories
Description
WinDaq Data Acquisition software unlock code includes a CD (shipping charges apply). Allows you to record at speeds up to 150kHz throughput.
Description
WinDaq Data Acquisition software unlock code on CD (shipping charges apply). Allows you to record at speeds up to 150kHz throughput and record different channels at different rates.
Description
Upgrade to the latest version of WinDaq/Pro+. For registered users of WinDaq/Pro+ ONLY. CD will be shipped - shipping charges apply.
Description
Upgrade to the latest version of WinDaq with the latest unlock code. For registered users of WinDaq/Pro+ ONLY. No CD will be shipped. Unlock code ONLY. Please allow at least one business day for processing.
Description
Upgrade to the latest version of WinDaq with the latest unlock code. For registered users of WinDaq/Pro ONLY. CD will be shipped - shipping charges apply.
Description
Upgrade to the latest version of WinDaq with the latest unlock code. For registered users of WinDaq/Pro ONLY. No CD will be shipped. Unlock code ONLY. Please allow at least one business day for processing.
Description
16-channel DI-8B module expansion backpack. Includes interconnecting cable, power cable, and a pair of stacking brackets to mount on top of your Data Acquisition System. Learn more.
Description
BLACK shrouded banana plugs - both ends - silicone.
Description
RED shrouded banana plugs - both ends - silicone.
Description
BLACK Right angle insulated plunger hook clip.
Description
RED Right angle insulated plunger hook clip.
Description
Spare 5-pin power connector for DI-718Bx, DI-720, DI-722, and DI-730 instruments.
Description
250 Ω ±0.1% Precision, Metal Film, 1/2W, Axial Resistor
Exceptionally low noise; typically 0.10 μV/V
Low voltage coefficient to ±5 ppm/V
Flame retardant coating. RoHS
compliant. Learn More
Description
Re-calibration Service, this calibration certificate is included with new instruments. NIST-traceable calibration certificate for all DI-730 data acquisition and data logger products. Shipping charges do apply when ordering online. Learn more about Calibrating your device.
Software
WinDaq
Data Acquisition Recording and Playback Software
Included FREE with your DAQ hardware purchase. Record up to 240 samples per second throughput. Installs drivers and WinDaq recording and playback software. Includes optional Add-on software with a valid installation key. Record higher sample rates with an Unlock Code (see Accessories).
ActiveX Controls
Programming
FREE ActiveX programming solutions for Dataq Instruments products. Free sample programs available. Compatible with LabView.
WinDaq/XL
WinDaq-to-Excel Real Time Link
Real-time data logging directly to Microsoft Excel (supports MS Office 2003 or higher). No programming required. Free Trial Version. Purchase full version on WinDaq/XL product page.
Advanced CODAS
$595
Advanced analysis software add-on to WinDaq Waveform Browser software. Allows calculated channels, derivative, integration, moving average, automatic file management, automatic scaling, and more.
Specifications
Analog Inputs | |||||||||||||||||||||||||||||
Number of Channels: | 8 wide range and 16SE/8DI general purpose | ||||||||||||||||||||||||||||
Input Type: | Bipolar | ||||||||||||||||||||||||||||
Isolation: | ±1000V input-to-output and channel-to-channel | ||||||||||||||||||||||||||||
Analog Resolution: | 14-bit, 1 part in 16,384 | ||||||||||||||||||||||||||||
Maximum normal mode voltage (VNM): | 1500VDC or peak AC | ||||||||||||||||||||||||||||
Maximum common mode voltage (VCM): | 1000VDC or peak AC | ||||||||||||||||||||||||||||
Max Sample Rate: | 150,000 Samples per second throughput | ||||||||||||||||||||||||||||
Measurement Range Full Scale: |
* Usable over +8159 to -8160 ADC counts
|
||||||||||||||||||||||||||||
Common Mode Rejection: | 100dB min @ DC to 60Hz | ||||||||||||||||||||||||||||
Accuracy: | Vin <e; 800V ±(0.25% of full scale range ±100µV) Vin > 800V ±(0.5% of full scale range) |
||||||||||||||||||||||||||||
Channel-to-Channel Crosstalk: | -75 dB @100kHz into 100 Ω unbalanced | ||||||||||||||||||||||||||||
Input Impedance: | 10MΩ for all ranges | ||||||||||||||||||||||||||||
Max jitter between synched units: | 5 microseconds | ||||||||||||||||||||||||||||
Analog frequency response (each channel): |
|
||||||||||||||||||||||||||||
Analog Outputs | |||||||||||||||||||||||||||||
Number of Channels | 1 | ||||||||||||||||||||||||||||
Resolution: | 12-bit, 1 part in 4096 @ 250kHz | ||||||||||||||||||||||||||||
Output Voltage Range: | ±10V | ||||||||||||||||||||||||||||
Output Impedance: | 10Ω | ||||||||||||||||||||||||||||
Sample Throughput Rate: | 40,000 standard; 80,000 bi-directional | ||||||||||||||||||||||||||||
Output Offset Voltage: | 1 bit max @ 1kHz sample rate | ||||||||||||||||||||||||||||
Gain Error: | 1 part in 16,384 @ 1kHz | ||||||||||||||||||||||||||||
Triggering | |||||||||||||||||||||||||||||
Pre-trigger length: | 64,000 samples | ||||||||||||||||||||||||||||
Post-trigger length: | 64,000 samples | ||||||||||||||||||||||||||||
Trigger channel: | any channel | ||||||||||||||||||||||||||||
Trigger level hysteresis: | 8-bit (256 counts) | ||||||||||||||||||||||||||||
Intelligent Oversampling | |||||||||||||||||||||||||||||
Modes: | Signal averaging, maximum value, minimum value, last point, frequency, and RMS | ||||||||||||||||||||||||||||
Physical/Environmental | |||||||||||||||||||||||||||||
Dimensions: | 7.29W × 9L × 1.52H inches | ||||||||||||||||||||||||||||
Operating Temperature: | 0 to 70°C | ||||||||||||||||||||||||||||
Storage Temperature: | -55 to 150°C | ||||||||||||||||||||||||||||
Humidity: | 0 - 90% non condensing | ||||||||||||||||||||||||||||
Weight | 5 lbs. | ||||||||||||||||||||||||||||
Software | |||||||||||||||||||||||||||||
Supported Applications: | ActiveX Controls; WinDaq; WinDaq/XL; WinDaq Waveform Browser | ||||||||||||||||||||||||||||
Analog Inputs | |||||||||||||||||||||||||||||
Connector Type: | Safety banana socket - 1 pair per channel (wide range channels) |
Articles
Solutions for Electromechanical Maintenance and Troubleshooting
Variable Speed Drive Measurements
Variable Speed Drive Measurements
WinDaq Comes through in the Clutch
Bridge Structural Monitoring System
Troubleshooting Hydraulic Systems
WinDaq Cuts Maintenance Costs in Steel Mill
Casting Off with WinDaq and the DI-730
Moon Roof Design and Qualification
Learn the Importance of Isolation In Four Easy Lessons
Enhancing Data Acquisition with Intelligent Over Sampling (IOS)
Intelligent Oversampling Methods Extend Measurement Flexibility
Frequency and True RMS Data Acquisition
The Art and Science of Reliability
Videos
Introducing the DI-730
The DI-730 features eight, high voltage channels with a full scale measurement range of ±10mV to ±1000V.
Distributed and Synchronized
Distributed Synchronous Ethernet data acquisition with DI-720/730 Systems.